PhysioD

Explorer PhysioD front end, 16-channel true differential

True Differential Recording

PhysioD

Providing the flexibility to record any electrophysiology signals, Ripple Neuromed’s PhysioD Front End can be used for EMG, ECG, respiration, or other non-invasive measures. Available in a true-differential 16-channel configuration, the PhysioD Front End can meet all of your high-density, peripheral recording needs.

Features

  • 16 channel differential pairs
  • Acquire EMG, ECG, respiration, and other bio-signal measures

Specifications

PhysioD Specifications

Compatible Processors Summit
Channels 16, true differential referencing
Bit Depth 24-bit
Resolution 0.022μV/bit
Sampling Rate 7.5kHz
Bandwidth DC to 2kHz
CMRR 110dB
Inferred Noise <1µVrms at 7.5kHz
Input Range Min ± 0.1875V
Input Range Max ± 2V
Input Impedance 1GΩ, 20pF
Back Explorer Summit

Next step

Connect to electrodes