ripple neuromed research tools

EEG and EEG+tES

Designed for non-invasive high and low impedance electrodes

macro-cropped

Macro and Macro+Stim

Designed for invasive low-impedance macroelectrodes

micro2-stim-sized

Micro2 and Micro2+Stim

Designed for invasive high-impedance microelectrodes

physiod-sized

Physio and PhysioD

Designed for both invasive and non-invasive EMG and peripheral recordings

Non-Invasive Recording and Stimulation

EEG and EEG+tES

Designed for both high and low impedance surface electrodes, Ripple Neuromed’s EEG Front Ends are suited for any single reference surface recording.

Integrate your standard touchproof electrodes, caps, or nets with any of our Front Ends. If your experiments call for stimulation, our EEG+tES Front End is versatile enough to deliver transcranial (tES), transcutaneous nerve (TENS), or spinal stimulation (TSS).

Features

  • Compatible with any electrode technology, including caps or nets
  • Wearable on a cap or net to achieve the signal quality of active electrodes with the durability and robustness of passive electrodes
  • Compact form factor is ideal for mobile and ambulatory EEG applications

Primary Use Cases

HD EEG | ERP | BCI | Mobile EEG | Sleep Research | Neuromodulation

ripple neuromed research tools

Start typing and press Enter to search

Front Ends Comparison Chart

Micro2 Macro EEG Physio
Recording
Channels 32 32 or 128 32 or 128 16 bipolar or 32 shared reference
Resolution 16-bit 24-bit 24-bit 24-bit
Max Input Range ± 12mV ± 2V ± 2V ± 2V
Sampling Rate Up to 30kHz Up to 7.5kHz Up to 7.5kHz Up to 7.5kHz
Bandwidth 0.1Hz – 7.5kHz DC – 2 kHz DC – 2kHz DC – 2kHz
(+Stim version) (1Hz – 7.5kHz) (0.03Hz – 2kHz) (DC – 2kHz) (N/A)
Stimulation
Version +Stim version +Stim version +tES version Not available
Stimulation Range (per channel) Up to ± 1.5mA Up to ± 15mA Up to ± 5mA N/A
Pulse Width Minimum 33µs 33µs 33µs N/A
Compliance Voltage ± 8.5V ± 9 to 30V (configurable) ± 9 to 30V (configurable) N/A

EEG Specifications

Specification Value
Compatible Processors Summit, Quest HD, and Quest HD+Stim
Channels 32 or 128, single reference
Bit Depth 24-bit
Resolution 0.022μV/bit
Sampling Rate 7.5kHz
Bandwidth DC to 2kHz
CMRR 110dB
Inferred Noise <1 µVrms at 7.5kHz
Input Range Min ± 0.1875V at a gain of 24
Input Range Max ± 2V
Input Impedance 1GΩ, 20pF

EEG + tES Specifications

Specification Value
Compatible Processors Summit or Quest HD+Stim
Channels 32 or 128, single reference
Bit Depth 24-bit
Resolution 0.022μV/bit
Sampling Rate 7.5kHz
Bandwidth DC to 2kHz
CMRR 110dB
Inferred Noise <1 µVrms at 7.5kHz
Input Range Min ± 0.1875V at a gain of 24
Input Range Max ± 2V
Input Impedance 1GΩ, 20pF
Stimulation Waveforms tDCS, tRNS, tACS, tPCS, user defined
Stimulation Current Range ±5mA per channel
Compliance Voltage ±9 to 30V (configurable)
Min Pulse Width 33µs
Stimulation Frequency DC to 15kHz